  • Materials
  • Protocol
  • Trouble Shooting
  • Image & Flash
     

Problems

Possible Sources

Solutions

No signal

Incorrect or no Detection Antibody was added

Add appropriate Detection Antibody and continue

Substrate solution was not added

Add substrate solution and continue

Incorrect storage condition

Check if the kit is stored at recommended condition and used before expiration date

     

Poor Standard Curve

Standard was incompletely reconstituted or was inappropriately stored

Aliquot reconstituted standard and store at  -70 ℃

Imprecise / inaccurate pipetting

Check / calibrate pipettes

Incubations done at inappropriate temperature, timing or agitation

Follow the general ELISA protocol

Background wells were contaminated

Avoid cross contamination by using the sealer appropriately

     

Poor detection value

The concentration of antigen in samples was too low

Enriching samples to increase the concentration of antigen

Samples were ineffective

Check if the samples are stored at cold environment. Detect samples in timely manner

     

High Background

Insufficient washes

Use multichannel pipettes without touching the reagents on the plate

Increase cycles of washes and soaking time between washes

TMB Substrate Solution was contaminated

TMB Substrate Solution should be clear and colorless prior to addition to wells

Materials were contaminated

Use clean plates, tubes and pipettes tips

     

Non-specificity

Samples were contaminated

Avoid cross contamination of samples

The concentration of samples was too high

Try higher dilution rate of samples

 

 

Human TNFRSF10D / TRAIL R4 ELISA Pair Set Related Products & Topics

Related Areas:

Immunology>>Cytokine & Receptor>>TNF Superfamily>>TRAIL R4/TNFRSF10D/CD264

Immunology>>Cluster of Differentiation>>B Cell CD Antigen>>TRAIL R4/TNFRSF10D/CD264

Immunology>>Cluster of Differentiation>>Granulocyte CD Antigen>>Neutrophils>>TRAIL R4/TNFRSF10D/CD264

Proteins:

Molecule Species Description //For Detailed Info. and Price------CLICK! Cat. No
TRAIL R4/TNFRSF10D/CD264 Human TRAIL R4/TNFESF10D/CD264/Fc Protein, Recombinant 10413-H03H
TRAIL R4/TNFRSF10D/CD264 Human TRAIL R4/TNFESF10D/CD264 Protein, Recombinant 10413-H08H

Antibodies:

Background

TNF-related apoptosis-inducing ligand (TRAIL), also known as Apo-2 ligand and TNFSF10, are TNF superfamily members, and preferentially induces apoptosis in tumor cell lines, but not in normal cells. TRAIL binds at least five different receptors with characteristic cysteine-rich repeats. Human TRAILR4, also known as decoy receptor 2 (DcR2) and TRUNND (TRAIL receptor with a truncated death domain), is a member of the TNF receptor superfamily and is referred to as TNFRSF10D. TRAILR4 shows a unique expression pattern in human tissues and is particularly abundant in fetal liver and adult testis. This protein is a type 1 transmembrane receptor consisting of an extracellular TRAIL-binding domain, a transmembrane domain, and a truncated cytoplamic death domain. Accordingly, TRAILR4 is not capable of activating TRAIL-mediated NF-kappaB or apoptosis pathways, and identified as an antiapoptotic receptor, although it retains one third of a consensus death domain motif in its cytoplasmic domain. It has been demonstrated that TRAILR4 expression is directly regulated by p53 and attenuates cell response to DNA-damaging agents.