Product Catalog



COQ7 ELISA Pair Set, Human ELISA Pair Set
| Catalog | Size (Price) | Quantity | In Stock | Operation | Other Information |
| SEK12195 |
|
YES |
|
![]() |
Human COQ7 ELISA Pair Set
|
|
Order or Inquire for COQ7 ELISA Pair Set |
|
|
High affinity Human COQ7 ELISA Pair Set |
|
|
Detection limit - 39.1 pg/ml |
|
|
Affordable price and 30%-80% cost saving for Bulk order |
- Materials
- Protocol
- Trouble Shooting
- Image & Flash
|
|
|
|
|
No signal |
Incorrect or no Detection Antibody was added |
Add appropriate Detection Antibody and continue |
|
Substrate solution was not added |
Add substrate solution and continue |
|
|
Incorrect storage condition |
Check if the kit is stored at recommended condition and used before expiration date |
|
|
Poor Standard Curve |
Standard was incompletely reconstituted or was inappropriately stored |
Aliquot reconstituted standard and store at -70 ℃ |
|
Imprecise / inaccurate pipetting |
Check / calibrate pipettes |
|
|
Incubations done at inappropriate temperature, timing or agitation |
Follow the general ELISA protocol |
|
|
Background wells were contaminated |
Avoid cross contamination by using the sealer appropriately |
|
|
Poor detection value |
The concentration of antigen in samples was too low |
Enriching samples to increase the concentration of antigen |
|
Samples were ineffective |
Check if the samples are stored at cold environment. Detect samples in timely manner |
|
|
High Background |
Insufficient washes |
Use multichannel pipettes without touching the reagents on the plate |
|
Increase cycles of washes and soaking time between washes |
||
|
TMB Substrate Solution was contaminated |
TMB Substrate Solution should be clear and colorless prior to addition to wells |
|
|
Materials were contaminated |
Use clean plates, tubes and pipettes tips |
|
|
Non-specificity |
Samples were contaminated |
Avoid cross contamination of samples |
|
The concentration of samples was too high |
Try higher dilution rate of samples |
Human COQ7 ELISA Pair Set Related Products & Topics
Related Areas:
Proteins:
Antibodies:
Background
Ubiquinone biosynthesis protein COQ7 homolog, also known as Coenzyme Q biosynthesis protein 7 homolog, Timing protein clk-1 homolog and COQ7, is a mitochondrion inner membrane and peripheral membrane protein which belongs to theCOQ7 family. It is expressed dominantly in heart and skeletal muscle. COQ7 is synthesized as a preprotein that is imported into the mitochondrial matrix, where the sequence is cleaved off and the mature protein becomes loosely associated with the inner membrane. This enzyme is responsible for the hydroxylation of 5-demethoxyubiquinone to 5-hydroxyubiquinone. Human COQ7 protein is mostly helical, and contains an alpha-helical membrane insertion. It has a potential N-glycosylation site, a phosphorylation site for protein kinase C and another for casein kinase II, and three N-myristoylation sites. COQ7 is involved in lifespan determination in ubiquinone-independent manner. It is also involved in ubiquinone biosynthesis. COQ7 is potential central metabolic regulator.

+86-400-890-9989







Problems
Possible Sources
Solutions 